Zero Air Generators
Parker Balston Zero Air Generators will produce up to 30 lpm of zero grade air.
-
-
-
-
- Parker Zero Air Generators are systems manufactured with state-of-the art, highly reliable components engineered for easy installation, operation, and long term performance. Parker Zero Air Generators are much easier to install than dangerous high pressure gas cylinders, and only need to be installed once! All that is required is a standard compressed air line and an electrical outlet.
-
-
-
- Parker Zero Air Generators are easy to operate, there is no complicated operating procedure to learn or any labor intensive monitoring required. Parker Zero Air Generators eliminate all the inconveniences and costs of cylinder gas supplies and dependence on outside vendors. Uncontrollable vendor price increases, contract negotiations, long term commitments, and tank rentals are no longer a concern. Parker Zero Air Generators offer long term cost stability.
-
-
-
- There is no need to use valuable laboratory floor space to store excessive reserves to protect yourself from late deliveries, transportation interruptions, or periods of tight supplies. With a Parker Zero Air Generator, you control your supply.
-
-
-
- Features/Benefits
-
-
-
- • Produce UHP Zero Air from house compressed air (<0.05 ppm THC)
-
-
-
- • Eliminate inconvenient and dangerous zero air cylinders from the laboratory
-
-
-
- • Increase the accuracy of analysis and reduce the cleaning requirement of the detector
-
-
-
- • Qualitative SMART-Display provides operational status at a glance
-
-
-
- • Recommended and used by many GC and column manufacturers
-
-
-
- • Typical payback period of less than 1 year
-
-
-
- • Silent operation and minimal operator attention required
-
-
-
- • Models available to service up to 66 FIDs
-
-
-
BRAND
Balston
APPLICATION
Analytical Laboratory
FOR FLUID TYPE
Zero Air
MAXIMUM INLET PRESSURE RATING
125 psig
GAS PURITY
75-83: <0.1 ppm Hydrocarbons, HPZA-3500: <0.05 ppm Hydrocarbons, HPZA-7000: <0.05 ppm Hydrocarbons, HPZA-18000: <0.05 ppm Hydrocarbons, HPZA-30000: <0.1ppm
FLOW RATE
75-83: 1 LPM, HPZA-3500: 3.5 LPM, HPZA-7000: 7 LPM, HPZA-18000: 18 LPM, HPZA-30000: 30 LPM
COMPRESSOR INCLUDED
No
INLET PORT SIZE
1/4 (NPT Female) inch
OUTLET PORT SIZE
1/4″ NPT Female
SPECIFICATIONS MET
IEC 1010-1, CSA, UL 3101, CE
ELECTRICAL REQUIREMENTS
120 VAC/60 Hz
MINIMUM INLET PRESSURE RATING
40 psig
PORT SIZE
Inlet: 1/4″ NPT, Outlet: 1/4″ NPT
Explosion-Proof Zero Air Generator
The Parker Balston expolosion proof zero air generator produces up to 1,000 cc/min of high purity zero grade air.
-
- The Parker explosion proof zero air generators are a complete system with carefully matched components engineered for easy installation, operation and long term reliability. They are designed to transform standard compressed air into a safe supply of hydrocarbon free (< 0.1ppm) zero grade air.
-
- The explosion proof zero air generators are an ideal alternative to high purity air cylinders used on process GC-FID’s in explosive environments. Simply connect to the GC-FID to ensure a low baseline signal to noise ratio resulting in consistent, accurate, and reliable analysis. Payback periods are typically less than one year.
-
- Features/Benefits
-
- • Produces a supply of high purity air for process GC-FID’s in explosive environments
-
- • Payback periods typically less than one year
-
- • Compact design mounts directly to Unistrut® framing
-
- • CENELEX Certification LCIE 03 ATEX 6232: Code EEX of 11B and H2 T6, Category 112G
-
- • CENELEX standard: EN 50014: 1997 +A1, A2, EN 50018: 2000
- • Compact, reliable, no moving parts, and minimal maintenance
APPLICATION
Analytical Laboratory
FOR FLUID TYPE
Zero Air
BRAND
Balston
MAXIMUM INLET PRESSURE RATING
125 psig
9 barg
MINIMUM INLET PRESSURE RATING
40 psig
3 barg
ELECTRICAL REQUIREMENTS
120 VAC/60 Hz, 0.5 amps
Zero Air Generators For GC Combustion Detector Applications
Ultra high purity, organic free, air for GC combustion detectors that will Increase resolution and detection limits of analysis. Flow rates range from 1 L/min to 30 L/min.
-
- The Parker domnick hunter UHP-ZA zero air generators produce a continuous stream of organic impurity free air from an external dry compressed air source and offer superior limits of detection over and above other modes of supply.
-
- The UHP-ZA generators feature an interchangeable top panel facilitating the direct mounting of any Parker domnick hunter hydrogen generator. The stackable system forms an innovative, modular FID gas station suitable for all known GC combustion detectors such as FID, FPD and NPD.
-
- UHP-ZA generators may also be used in many other chemical analysis and life science applications, including LC/MS source gas, zero and combustion gas for total hydrocarbon analysers and as a gas sensing calibration and dilution gas.
-
- Features / Benefits
-
- • Ultra high purity, organic free, air for GC combustion detectors
-
- • Increase resolution and detection limits of analysis
-
- • Compact, reliable with minimal operator attention and maintenance
-
- • Eliminate inconvenient and potentially dangerous air cylinders
-
- • Payback period typically less than 24 months
- • Models available to supply up to 75 FID’s
GAS TYPE GENERATED
Zero Air
AIR CONSUMPTION
1.2 to 35 L/min
MAXIMUM OUTLET PRESSURE RATING
10 barg
145 psig
MINIMUM OPERATING TEMPERATURE
5 °C
40 °F
MAXIMUM OPERATING TEMPERATURE
40 °C
104 °F
INLET PORT SIZE
1/8 or 1/4 (Compression Fitting) inch
OUTLET PORT SIZE
1/8 or 1/4 (Compression Fitting) inch
INLET AIR QUALITY SPECIFICATIONS
ISO8573-1 Class 3.2.1
HEIGHT
325 to 455 mm
12.8 to 17.9 inch
WIDTH
340 mm
13.4 inch
DEPTH
425 mm
16.7 inch
WEIGHT
10.2 to 15.2 kg
22.5 to 33.5 lb
PREVENTIVE MAINTENANCE KIT
12 month inlet filter kit: 005AO
’12 month outlet filter kit: 005AA
24 month fan pm kit: 606272525 or 12 month inlet filter kit: 005AO
’12 month outlet filter kit: 005AA
24 month fan pm kit: 606272526
Literature and Reference Materials
FOCUSED ON Continuous Generation – Nitrogen gas generation solutions for liquid chromatography mass spectrometry – (English)
Catalogue – Parker domnick hunter Analytical Gas Systems – (English)
Catalogue – Parker domnick hunter Gas Chromatography Products – (English)
FOCUSED ON Exceptional Accuracy – Gas generation solutions for gas chromatography – (English)
Product Information Sheet – Zero Air Generators for GC Combustion Detector Applications – (English)
White Paper – An evaluation of the analysis of polyaromatic hydrocarbons (PAHs) using a gas chromatograph-flame ionisation detector (GC-FID) with generated & cylinder/compressor gases.
Product Demos & Tutorials
Interactive Presentation – Parker domnick hunter Analytical Gas Systems – English (Requires Adobe Flash)
High Purity Nitrogen And Dry Air Generators
Ideal for analytical instruments that require high purity nitrogen and air. Produces a continuous supply of high purity nitrogen (99.999%) and dry air.
Parker domnick hunter G6 and G7 nitrogen and dry air generators employ robust, field proven technology to produce ultra high purity nitrogen and dry air suitable for chemical analysis applications.
The G6 and G7 generators provide a continuous stream of nitrogen and dry air from a single ‘plug & play’ unit, are extremely quiet in operation and are fully approved for use by major instrumentation manufacturers.
Innovative technology, design and function combine to completely eliminate all other modes of supply, facilitating maximum instrument uptime, attractive return on investment and proven analytical performance.
Features / Benefits
• Ideal for analytical instruments that require high purity nitrogen and air
• Produces a continuous supply of high purity nitrogen 99.999% and dry air
• Integral oil free compressor
• Eliminate inconvenient and potentially dangerous nitrogen cylinders
• Compact, reliable with minimal operator attention and maintenance
• Improve analysis and reproducibility with guaranteed high purity gas
GAS TYPE GENERATED
Nitrogen and Dry Air
GAS PURITY
Nitrogen: >99.999%
Dry Air: -40°C (dew point)
OUTLET PORT SIZE
1/8 or 1/4 (Compression Fitting) inch
MAXIMUM OUTLET PRESSURE RATING
5 barg
72.5 psig
INTEGRAL COMPRESSOR INCLUDED
Yes
MINIMUM OPERATING TEMPERATURE
5 °C
45 °F
MAXIMUM OPERATING TEMPERATURE
41 °C
113 °F
ELECTRICAL REQUIREMENTS
207V – 253V 50/60Hz
HEIGHT
842 or 874 mm
33.1 or 34.4 inch
WIDTH
345 mm
13.6 inch
DEPTH
413 or 663 mm
16.3 or 26.1 inch
WEIGHT
58 or 93 kg
127.9 or 205 lb
PREVENTIVE MAINTENANCE KIT
12 month filter kit: 606272351
12 month compressor kit: 606272336 or 12 month filter kit: 606272351
12 month compressor kit: 606272337 or 12 month filter kit: 606272352
12 month compressor kit: 606272334 or 12 month filter kit: 606272352
12 month compress
Literature and Reference Materials
Catalogue – Parker domnick hunter Analytical Gas Systems – (English)
Parker–AGS Supplies Catalog | AGS-SUPCAT-M
Product Information Sheet – High Purity Nitrogen and Dry Air Generators – (English)
Product Demos & Tutorials
Interactive Presentation – Parker domnick hunter Analytical Gas Systems – English (Requires Adobe Flash)
Zero Air Generators (Low Flow Up To 1 L/Min)
Extremely reliable zero air generators providing air free of total hydrocarbons to <0.1 ppm at flow rates up to 1 L/min.
-
- The Parker Balston Zero Air Generator can produce up to 1 L/min of high purity zero grade air. Compressed air is pre-filtered down to 0.01 micron and then purified using a state of the art combined heated catalyst module.
-
- There are no moving parts and no noise, making the generator extremely reliable and ideal to install in the laboratory. Simple and quick to install, the Zero Air Generator requires minimal maintenance just once a year.
-
- The resultant air is free of total hydrocarbons (THC) to < 0.1ppm making it ideal for all FID applications. The low levels guarantee a low signal to noise ratio, ensuring a flat constant base line with no peaks or fluctuations.
-
- Features / Benefits
-
- • Ultra high purity air for GC / FID applications
-
- • Payback period typically less than one year
-
- • Silent operation and minimal operator attention required
-
- • Eliminate inconvenient and potentially dangerous air cylinders from the laboratory
-
- • Models available to service up to 75 FIDs
- • Increases the accuracy and repeatability of analysis
FLOW RATE
Up to 1 L/min
GAS TYPE GENERATED
Zero Air
GAS PURITY
< 0.1ppm THC
MAXIMUM OPERATING TEMPERATURE
35 °C
95 °F
MINIMUM OPERATING TEMPERATURE
10 °C
50 °F
MAXIMUM OUTLET PRESSURE RATING
8 barg
116 psig
ELECTRICAL REQUIREMENTS
230v – 50Hz
OUTLET PORT SIZE
1/4 (NPT Female) inch
HEIGHT
250 mm
9.85 inch
WIDTH
300 mm
11.8 inch
DEPTH
80 mm
3.15 inch
WEIGHT
2 kg
4.4 lb
PREVENTIVE MAINTENANCE KIT
12 Months – MK7583
To request further information: